Specifications
Technical data | |
---|---|
Measuring quantity | Standard flow velocity wN based on standard conditions 1) Temperature of medium TM (optional) |
Measured medium | Air or nitrogen, other gases upon request |
Measuring range wN | 0 … 2.5 / 10 / 20 / 40 / 50 / 60 m/s |
Lower detection limit wN | 0.2 m/s |
Measuring range TM (optional) | -20 … +120 °C |
Measuring accuracy | |
Standard wN | ±(5 % of measured value + [0.4 % of fmr; min. 0.02 m/s]) 2) |
High precision (optional) wN | ±(3 % of measured value + [0.4 % of fmr; min. 0.02 m/s]) 2) |
Response time (t90) wN | 3 s (jump from 0 to 5 m/s) |
Temperature gradient wN | < 8 K/min (at wN = 5 m/s in air) |
TM (wN > 2 m/s) | ±1 K (0 … 40 °C) ±2 K (remaining measuring range) |
Operating temperature | |
Sensor probe | -20 … +120 °C |
Electronics | 0 … +70 °C |
Storage temperature | -20 … +85 °C |
Material | |
Sensor head | Platinum element (glass passivated); PPO/PA |
Sensor tube | Stainless steel 1.4571 |
Enclosure | PBT (glass-fibre reinforced) |
Sensor cable | PVC |
General data | |
Humidity | Measuring mode: Non-condensing (< 95 % RH) |
Operating pressure (max.) | Atmospheric (700 … 1,300 hPa) |
Display | LED green: Operating status LED red (flashing): Sensor defective |
Supply voltage | 24 V DC ± 10 % |
Current consumption | < 60 mA |
Analog output | Short circuit protected; load capacity CL ≤ 10 nF |
- Standard: wN | 1 x 0 … 10 V (RL ≥ 10 kΩ) or 4 … 20 mA (RL ≤ 300 Ω) |
- Option: wN and TM | 2 x 4 … 20 mA (RL ≤ 300 Ω) |
Electrical connection | Non-detachable cable, , 4 x 0,14 mm2, with ferrules: - Standard: Length 2 m - Special length: 3 … 100 m (step: 1 m) |
Maximum cable length | Voltage interface: 15 m; current interface: 100 m |
Mounting position | Arbitrary |
Mounting tolerance | ±3° relative to flow direction (unidirectional) |
Protection type | Sensor probe: IP66; enclosure: IP54 |
Protection class | III (SELV) or PELV (EN 50178) |
Probe length | 50/100/200/350/500 mm |
Weight | 200 g max. |
1) Standard conditions: TN = 20 °C and pN = 1,013.25 hPa
2) Related to adjustment reference; "fmr" = final measuring range
Subject to technical changes.