Specifications
Technical data | |
---|---|
Measuring quantity | Standard flow velocity wN based on standard conditions 1) Temperature of medium TM |
Measured medium | Clean a or nitrogen, additional gases on request |
Measuring range wN | 0 … 1 / 2.5 / 10 m/s |
Max. display range wN | +10 % above measuring range |
Lower detection limit wN | 0.1 m/s |
Measuring range TM | 0 … +60 °C |
Measuring accuracy | |
Standard wN | ±(3 % of measured value + 0.05 m/s) 2) |
High precision (optional) wN | ±(1 % of measured value + 0.04 m/s) 2) |
Response time (t90) wN | 15 s |
Temperatur TM (wN > 0.3 m/s) | ±1 K (10 ... 30 °C); ±2 K (remaining measuring range) |
Operating temperature | |
Sensor | 0 … 60 °C |
Storage temperature | -20 … +85 °C |
Material | |
Sensor head | Stainless steel 1.4404 |
Sensor tube | Stainless steel 1.4404 |
Plug connector, nut screw | Stainless steel 1.4404 |
Light ring | PC |
General data | |
Humidity | Measuring mode: Non-condensing (< 95 % RH) |
Operating pressure (max.) | Atmospheric (700 … 1,300 hPa) |
Supply voltage | 24 V DC ± 10 % |
Current consumption | tbd |
Analog outputs wN & TM | 0 … 10 V (RL ≥ 10 kΩ; short-circuit protected) 4 … 20 mA (RL ≤ 300 Ω>br> CL ≤ 10 nF |
Error signal | Only with current interface: 2 mA (according to NAMUR NE43) |
Electrical connection | Plug-in-connector M9, male, 7-pin (shielded), screwed |
Maximum cable length | Voltage interface: 10 m; current interface: 100 m |
Mounting position | Straight: ±3° relative to flow direction (unidirectional) Angled: Vertical downdraft flow (per design |
Protection type | IP65; electrical connector: IP67 (with correctly attached cable) |
Protection class | III (SELV) or PELV (EN 50178) |
Sensor length | Angled: 150/270 mm (vertical) x 300 mm (horizontal) Straight: 300 mm |
Weight | Approx. 280 g (angled design with 270 mm x 300 mm) |
1) Standard conditions: TN = 20 °C and pN = 1,013.25 hPa
2) Related to adjustment reference; "fmr" = final measuring range
Subject to technical changes.